.. inlab In-lab Test =================== For the in-lab test, only the main FTS part is used and the output is directedly measured by **Power Meter** or low-temperature millimeter wave dectector for RF source and black-body source, respectively. In this sense, the following steps are required to perform the in-lab test: .. note:: Follow necessary steps in :ref:`Hardware Preparation` before starting the in-lab test. - Open the application ``FTS.exe``. - Select and set ``SM Controller``. - Select ``Stepping`` mode. - Set ``Loop Number``. - Set ``Min Pos`` and ``Max Pos`` to adjust the spectra resolution. - Set ``Step`` and ``Pause Time``. - Confirm ``Linear Encoder Name`` is valid and with ``ctr0`` selected. - Make sure ``On Site Switch`` is switched off. - Leave ``Trigger Source`` empty. - Select and set the chopper signal in the first element of ``DAQmx Channels`` (``Saving Name`` can be **chopper**). - Select and set the detector signal in the second element of ``DAQmx Channels`` (``Saving Name`` can be **detector**). - Fill ``Data File Suffix`` and ``Comments`` on needs. - Click ``Start``.